84
GMC-I Messtechnik GmbH
Abbreviated commands:
Commands can be abbreviated with letters shown in boldface.
Letters not printed in boldface can be omitted.
Example: "
OU
TPUT ON" = "OU ON"
As a rule, letters can be entered in upper or lower case.
Stringing commands together:
Several commands in a single data string must be separated by semicolons (;).
Example: "USET 12; ISET 8.5; OUTPUT ON"
Formats for numeric parameters:
n :
Whole number (integer)
v :
Whole number, fixed or floating decimal point number with or without exponent
Examples: “12.5”, “0012.5”, “1.25E1”, “+1.25 e+01”
Function-Specific Comman
ds and Device Settings
SS
Et
txt
Set a message signal (sequence / analog interface)
X
X
X
X
X oFF
txt:
oFF, on
(message signal off/on)
StoP
n
Stop address for the sequence
X
X
11
n:
11; ... ; 255
Strt
n
Start address for the sequence
X
X
11
n:
11; ... ; 255
StA
RT_STOP
n1, n2 n:
11; ... ; 255
X
X
X
11; 11
STO
RE
n, v1, v2, v3, txt
Direct transfer of parameter data to memory
X
X
no effect
n:
11; 12; ... ; 253; 254; 255, memory address
v1:
Voltage setpoint
v2:
Current setpoint
v3:
Memory location-specific dwell time
txt:
Switching function, ON or OFF signal level, CLR: complete deletion of memory location data
td
EF
v
Dwell time independent of memory location (sequence)
X
X
X
X
0.01 s
v:
00.01; ... , 99.99 (in seconds)
trG
txt
Function selection for the trigger input (analog interface)
X
unchanged
T_M
ODE
txt
txt:
oFF
(trigger input deactivated)
X
X
out
(trigger input influences the OUTPUT)
rcL
(memory recall, step-by-step)
SEq
(sequence execution control)
LLO
(control panel disabling control)
UI
(Min-Max measured value storage control)
remote: “MIN”
tS
Et
v
Memory location-specific dwell time (sequence)
X
X
X
X
X tdef
v:
00.00; 00.01; ... , 99.99 (in seconds), at 00.00: value from tdEF automatically
UI
txt
Min-Max storage for measured U and I values (setup)
X
X
oFF
MI
NMAX
txt
txt
:
oFF; on; rSt
(Min-Max value storage off/on/reset)
X
X
X
UL
IM
v
Voltage setting limit (setup)
0
v
U
nom
X
X
X
X
U
nom
v:
15.0
(example: 15.0 V)
US
ET
v
Voltage setpoint
0
v
ULIM
X
X
X
X
X 0.000
v:
12.5
(example: 12.5 V)
W
AIT
v
Additional waiting time
X
X
v:
0.001;0.002 ... 9,999 (waiting time in seconds)
Gener
a
l Commands and Settings
*
CLS
Clear status command
X
X
no effect
DCL
,
SDC
Clear input and output buffers, all settings and register contents unchanged
X
X
no effect
*
DDT
txt
Define device trigger
X
X
DDT memory
txt:
max. 80 characters, example: *DDT USET 10/ISET 3/OUT ON
2)
cleared
ERAE
n
Device dependent event register A enable command
X
X
no effect
ERBE
n
Device dependent event register B enable command
X
X
no effect
*
ESE
n
Standard event enable command
X
X
no effect
IFC
INTERFACE CLEAR: Rest and initialize the IEC bus interface
(all settings and register contents remain unchanged)
X
X
*
OPC
Operation complete command
X
X
no effect
*
PRE
n
Parallel poll enable register command
X
X
no effect
*
PSC
n
Power-on status clear command
X
X
no effect
*
RCL
n
Recall and set
X
X
X
no effect
n:
1 ... 10
a stored device setting
11 ... 255 stored parameters: USET, ISET TSET and SSET
*
RST
Reset device to default settings
X
X
X
*
SAV
n
n:
0
*SAV 0 deletes the contents of memory locations
n1
to
n2
by using the address
parameters of the sequence function defined by START_STOP
n1
,
n2.
X
X
X
no effect
1 ... 10
Save current device settings
11 ... 253 Save momentary settings for USET, ISET TSET and SSET
254, 255 Reference values USET and ISET
*
SRE
n
Service request enable command
X
X
no effect
SRQ
Service request
X
no effect
*
TRG
Trigger: trigger the function specified with *DDT
X
X
no effect
*
WAI
Wait to continue command
X
X
no effect
Setting saved with *SAV n:
Default setting after
RESET *RST:
Expla-
nation in
Chapter
on Page
Manual setting or via interface
n=1
...
10
n=11
...
255
Setting Command Significance / Effect
Manual
IE
EE 488
RS 23
2C
Summary of Contents for SSP 120 Series
Page 2: ...2 GMC I Messtechnik GmbH ...