68
GMC-I Messtechnik GmbH
If necessary, the numeric value is rounded
off once again for the 4-place digital
display.
Default setting:
0.00 [A] (after RESET, *RST)
Example
(HP BASIC):
OUTPUT 713;"ISET 11.3"
{current setpoint: 11.3 A}
Comment
ISET may not be set to a value of greater than ILIM.
Any setting command to this effect is not executed.
Bit 1
would otherwise be set in event register B (
limit error
), and
bit 4
in the standard event register (
execution error
).
Query
Displays the selected output current setpoint ISET.
Programming
Query command:
IS
ET
?
Response string:
ISET
v
Length: 13
characters,
fixed
Parameter:
v (value)
Format: +nn.nnnn
Example
(HP Basic):
OUTPUT 713; "ISET?"
ENTER 713; A$
DISP A$
Display:
ISET +03.0000
MINMAX, MINMAX? – Min-Max Storage for Measured U and I
Values
Function and manual operation:
See explanation on page 27.
Settings
Programming
Setting command:
MI
NMAX
txt
Parameter:
txt (
status)
Parameter type:
Text (character)
Setting range
OFF
Storage of Min-Max values is
deactivated
ON
Storage of Min-Max values is
activated
RST
Contents of Min-Max memory are
reset, i.e. are replaced with the
momentary measured value for the
corresponding parameter.
Default setting:
OFF
(after RESET, *RST)
Example
(HP Basic):
OUTPUT 713;"MINMAX RST"
Query
Displays Min-Max memory enable status.
Programming
Query command:
MI
NMAX
?
Response string:
MINMAX txt
Length: 10
characters,
fixed
Parameter:
txt (status)
Example
(HP Basic):
OUTPUT 713; "MINMAX?"
ENTER 713; A$
DISP A$
Display:
MINMAX OFF
MODE? – Querying the Momentary Control Mode
Function and manual operation:
Displays the momentary output control mode.
Evaluation of condition register
CRA
.
Direct querying of the momentary control mode
See explanation in chapter 3 [3].
Programming
Query command:
MO
DE
?
Response string:
MODE txt
Length: 8
characters,
fixed
Parameter:
txt (control mode)
Query range:
OFF
output deactivated
CV
constant voltage regulation
CC
constant current regulation
OL
overload (power limiting)
Example
(HP Basic):
OUTPUT 713; "MODE?"
ENTER 713; A$
DISP A$
Display:
MODE CV
OCP, OCP? – Overcurrent Protection
Function and manual operation:
See explanation on page 26.
Settings
Programming
Setting command:
OC
P
txt
Parameter:
txt (
status)
Parameter type:
Text (character)
Setting range
OFF
Function inactive, continuous
current limiting (current regulation).
ON
Function activated, the output is
deactivated as soon as the
specified DELAY period has
elapsed.
Default setting:
OFF
(after RESET, *RST)
Example
(HP Basic):
OUTPUT 713;"OCP ON"
{activate OCP function}
Query
Displays the selected status of the current limiting function.
Programming
Query command:
OC
P
?
Response string:
OCP
txt
Length: 7
characters,
fixed
Parameter:
txt (status)
Example
(HP Basic):
OUTPUT 713; "OCP?"
ENTER 713; A$
DISP A$
Display:
OCP ON
Summary of Contents for SSP 120 Series
Page 2: ...2 GMC I Messtechnik GmbH ...