5.8 Trigger Input
Functions
The
TRG IN
floating
optocoupler input
allows for
remote control
of a
device function using a binary signal.
The function to be controlled is selected with the
T-MODE
setting (in the
trG
display).
☞
A detailed description is included on page 28 and page 74.
Connection
Connect the control signal between
TRG IN+
and
TRG IN-
. Refer
to the table below for the respective signal level.
The
TRIGGER
input can be driven with the
+15 V output
at the
analog interface via any desired switch (Figure 5.8).
Warning!
The TRG IN trigger input is a floating input and is
functionally isolated from the output current circuit.
This functional isolation is not equivalent to “safety
separation” as specified in electrical safety regulations.
Note
The trigger input is scanned roughly every 10 ms by the digital
control unit. After a signal change has been detected, repeated
querying ensues at short time intervals (suppression of switch
bouncing and interference pulses). This means that:
– Trigger signal pulses must have a minimum duration of
14 ms in order to assure reliable recognition.
– A delay of 1 to 15 ms may occur between application of the
control signal and triggering of the controlled function.
Figure 5.8
Controlling the Trigger Input with a Switching Element /
External Signal
5.9 Parallel Connection
If output current from a single KONSTANTER is insufficient for the
respective application, the outputs of
any number
of
KONSTANTERs can be parallel connected.
Attention!
!
If outputs with different nominal voltages are parallel
connected, all outputs must be limited to the lowest
utilized nominal voltage value. The ULIM parameter is
used to select this setting.
5.9.1
Direct Parallel Connection
Functions
Easiest way to provide the power consumer with
more current
than is available from a single KONSTANTER
KONSTANTERs with
differing nominal output voltages
can be
used. However,
all voltage setpoints
must be
set or limited to the
same value
.
Less suitable for
constant voltage regulating mode
Connection
Figure 5.9.1 a
Wiring for Direct Parallel Connection
Settings
Deactivate
all outputs.
Adjust
voltage setpoint USET
of all parallel connected
KONSTANTERs
to approximately the same value
:
Uset = USET1 = USET2 = USET3 = USETn
Adjust current setpoints ISET
such that they add up to the desired
cumulative current value Iset
:
Iset
= ISET1 + ISET2 + ISET3 + ... + ISETn
Activate
the outputs.
Functional Principle
After switching the outputs on,
load current
is initially supplied
by the
KONSTANTER with the highest voltage setting
.
If
load resistance
is continuously
reduced
, load current is
continuously increased.
When
load current
reaches the
ISET value
selected for the output
which is momentarily supplying power to the consumer,
current
limiting is activated for this output
.
If load resistance is further decreased,
current regulation reduces
output voltage until the voltage value of the
output with the next
lowest setting
is reached.
As of this point in time, this KONSTANTER also supplies a
portion of the load current
.
Output
Signal
U
s
I
s
OFF
High
4 ... 26 V DC
(U
s
2 V) / 1.5 k
ON
Low
0 ... 1 V DC
0 mA
+ SENSE
OUT
OUT
SENSE
Analog Interface
SIG1 OUT
SIG2 OUT
TRG IN
TRG IN
+15 V
AGND
Uset
Uset +
Iset +
U-MON
I-MON
+ OUT
+ OUT
Output
SSP KONSTANTER
SSP KONSTANTER
+ SENSE
OUT
OUT
SENSE
Analog Interface
SIG1 OUT
SIG2 OUT
TRG IN
TRG IN
+15 V
AGND
Uset
Uset +
Iset +
U-MON
I-MON
+ OUT
+ OUT
Output
Settings
USET = Uset
ISET = Iset
T_MODE
OFF
Settings
USET = Uset
ISET = Iset
T_MODE
OFF
Uout
Load
Uout
Iout
Load
Iout
I
s
U
s
I
s
approx. 10 mA
+ SENSE
OUT
OUT
SENSE
Analog Interface
SIG1 OUT
SIG2 OUT
TRG IN
TRG IN
+15 V
AGND
Uset
Uset +
Iset +
U-MON
I-MON
+ OUT
+ OUT
Output
Settings
USET = Uset
ISET1+2+3 = Iset
OUTPUT on/off
+ SENSE
OUT
OUT
SENSE
Analog Interface
SIG1 OUT
SIG2 OUT
TRG IN
TRG IN
+15 V
AGND
Uset
Uset +
Iset +
U-MON
I-MON
+ OUT
+ OUT
Output
Settings
USET = Uset
ISET1+2+3 = Iset
OUTPUT on/off
+ SENSE
OUT
OUT
SENSE
Analog Interface
SIG1 OUT
SIG2 OUT
TRG IN
TRG IN
+15 V
AGND
Uset
Uset +
Iset +
U-MON
I-MON
+ OUT
+ OUT
Output
Settings
USET = Uset
ISET1+2+3 = Iset
OUTPUT on/off
Load
= Only required for sensing mode operation
SSP KONSTANTER
SSP KONSTANTER
SSP KONSTANTER
Summary of Contents for SSP 120 Series
Page 2: ...2 GMC I Messtechnik GmbH ...