TDEF, TDEF? – Default Time for SEQUENCE Function
Functions:
See explanation on page 32.
Programming
Setting command:
TD
EF
v
Parameter:
v (value)
Parameter type:
Real
Setting range:
00.01
[s] –
99.99
[s]
Resolution:
10 ms
Default setting:
after RESET (*RST): unchanged
Example
(HP Basic):
OUTPUT 713;"TDEF 12.34"
Query
Recall the default value for standard dwell time.
Programming
Query command:
TD
EF
?
Response string:
TDEF
v
Length: 10
characters,
fixed
Parameter:
v (value)
Example
(HP Basic):
OUTPUT 713; "TDEF?"
ENTER 713; A$
DISP A$
Display:
TDEF 12.34
T_MODE, T_MODE? – Trigger Input Function Selection
Functions
The status of the
T_MODE
function is
not
saved as a device
setting with the
SAVE
key.
See explanations on page 28 -
trG.
Programming
Setting command:
T_M
ODE
txt
Parameter:
txt
Parameter type:
Action (character)
Setting range:
OFF
Trigger output deactivated
OUT
Trigger input influences
OUTPUT
:
output on/off
RCL
RECALL
: memory recall (step-by-
step)
1
SEQ
SEQUENCE
: sequential memory
recall (
SEQUENCE GO
)
1
LLO
LOCAL LOCKED
: Control panel
disabling; indicated by LOCKED /
SEQ LED at the front panel
MIN
MINMAX
: Min-Max memory control
for measured values
Default setting:
Unchanged (after RESET, *RST)
Example
(HP Basic)
:
OUTPUT 713;"T_MODE LLO"
{trigger input influences control panel
disabling}
1 If the SEQUENCE function has not first been initialized, an
active trigger signal results in error message “Err 27”, which
is displayed for approximately 1 second, and bit 5 (SEQE) is
set in event register B (ERB). “RCL” or “SEQ” settings can
only be executed if the SEQUENCE function has been fully
initialized.
Query
Displays selected trigger input function.
Programming
Query command:
T_M
ODE
?
Response string:
T_MODE
txt
Length: 10
characters,
fixed
Parameter:
txt (action)
Example
(HP Basic):
OUTPUT 713; "T_MODE?"
ENTER 713; A$
DISP A$
Display:
T_MODE LLO
TSET, TSET? Dwell Time Specific to Memory Location for the
SEQUENCE Function
Function and manual operation:
See explanations on page 31.
Setting value
TSET
is saved to
SETUP
memory as a device
setting with the
SAVE
command.
Programming
Setting command:
TS
ET
value
Parameter type:
Real
Setting range:
00.01
[s] ...
99.99
[s]
Resolution:
10 ms
Format:
nn.nn
Default setting:
after RESET (*RST): 00.00 = TdEF
Example
(HP Basic):
OUTPUT 713;"TSET 00.02"
Query
Recall memory location-specific dwell time.
Programming
Query command:
TS
ET
?
Response string:
TSET
v
Length: 10
characters,
fixed
Parameter:
v (value)
Format:
nn.nn
Example
(HP Basic):
OUTPUT 713; "TDEF?"
ENTER 713; A$
DISP A$
Display:
TDEF 00.02