4.9.4
Inserting a Memory Location
The start and stop addresses define the working range for the
command described below. Memory locations which do not lie
within this range are protected against access when the
command is executed. Each time a memory location is inserted,
the content of the previous stop address is lost.
Functions
•
Starting with the target address, previously saved data are
shifted to higher address numbers.
•
Memory location content located at memory address N prior
to insertion is shifted to memory address N+1.
•
Data which were previously stored to the memory location
with the stop address are lost.
•
The target address has no content after initial insertion.
Attention:
Currently selected start and stop addresses and start
and stop addresses saved to setup memory are not corrected
automatically!
Settings
☞
Select a start address and a stop address with the help of the
sequence submenu.
☞
Press the
<SAVE> key.
!
Sto
and the last selected address appear at the display.
☞
Select the desired target address with the <
> or the <
>
key.
☞
Simultaneously press the
<SAVE>
and <
> keys.
!
An
i
(for insert) appears in the right-hand display next to the
selected memory address number.
☞
Execute the insert command by pressing the <
ENTER
> key.
!
An empty memory location is added at the selected memory
address.
!
The address numbers of all memory locations above the
selected address are increased by 1.
Figure 4.9.4
Inserting a Memory Location Within a Given Range
CE/LOCAL
FUNCTION
RCL
Uout/Iout
Uset/Iset
FUNCTION
MENU
RCL
MENU
ENTER
SAVE
SAVE
&
Start and stop address
Abort
All memory locations
address 103 are shifted up by 1.
Memory location 103 is empty and the content of the stop address is lost.
Example of inserting an empty memory location within a given memory range:
01
10
occupied
011
255
occupied
occupied
empty
empty
start °1
sequence °1
stop °1
start °2
sequence °2
stop °2
curre
nt
s
ta
rt
curre
nt
s
top
S | t | o | Y | Y | Y | i
YYY
The previous content of the stop address is lost!
Insertion
YYY+
1
St
op
-1
St
op