This procedure is continued until load current triggers current
regulating at the output with the lowest voltage setting when
the
setpoint value for cumulative current
is reached.
This output maintains
constant
load current until the load
resistor is short-circuited.
Figure 5.9.1 b
U/I Diagram for Direct Parallel Connection
Notes:
Slightly
varying voltages
occur at the individual outputs as a
result of setting tolerances.
In the event of
larger voltage differences
, an
electronic sink
is
activated at the outputs with lower voltage settings.
The sink attempts –
in some cases in pulse mode
– to reach the
lower voltage value.
Neither the KONSTANTERs nor the power consumer are
damaged as a result.
If problems occur with
measurement of load current
, the
KONSTANTERs should be linked by means of
master-slave
parallel connection
The
outputs
can be activated and deactivated
commonly
by
connecting the
SIG1
outputs to the
TRG
inputs (see also chapter
Ideal working
current
Ideal working range
for voltage regulation
at the load
U
out1
U
out2
U
out3
I
out1
I
out2
I
out3
R
L
R
L
U
A
/
V
I
A
/
A
range for
regulation at
the load