90
GMC-I Messtechnik GmbH
8.7
System Messages
Error messages may appear at the digital display immediately
after the device is switched on, or after manual or remote control
triggering of certain functions.
Code
Meaning / Cause
Remedy
Err 1
ROM checksum error
ROM memory test failed
The device must be tested at a service
center and repaired if necessary.
Err 2
RAM write/read error.
RAM memory test failed
The device must be tested at a service
center and repaired if necessary.
Err 4
Write/read error.
Initialization of the IEC bus interface
failed:
– Electrical isolation, optocoupler
– Auxiliary power
– Controller
The device must be tested at a service
center and repaired if necessary.
Err 11
After execution of the self-test
(manual).
The measured value of the
manipulated parameter was
compared with the numeric setpoint
at the momentary operating point with
activated output and a deviation of
> n
resolution was detected (n is
defined internally, U: 15 LSD; I: 30
LSD).
➭
CAL
The device must be tested at a service
center and repaired if necessary.
Err 3
An unspecified self-test sub-function
has failed.
The device must be tested at a service
center and repaired if necessary.
Err 7
Serial interface disabled,
impermissible configuration, e.g. 8DB,
PE, 2SB, or 7DB, PN, 1SB. The error
message is displayed during the
power-up routine, or under the “BUS”
menu item after each entry.
Err 20
Attempted deletion and insertion
outside of the defined memory range
(from start to stop address).
Err 21
Error message during SEQUENCE
function operation, or after RECALL:
The (next) voltage or current setpoint
value to be recalled from the
SEQUENCE memory is higher than the
respective limit value (USET > ULIM or
ISET > ILIM). For this reason, memory
recall cannot be executed. The
SEQUENCE run is aborted.
Check the contents of the SEQUENCE
to be executed and coordinate
setpoints and limit values.
Err 22
Error message after SEQUENCE GO:
No executable values exist within the
memory range defined by the START
and STOP addresses for the
SEQUENCE run. The SEQUENCE run
cannot be started.
Check the selected START and STOP
addresses, as well as the contents
(USET, ISET and TSET) of the memory
locations defined by these
parameters.
Err 24
Attempted recall of an invalid value
(remote: *RCL xxx, where 011
xxx
255).
Err 25
Error message after OUTPUT ON:
Activation of the output is disabled by
an OUTPUT OFF signal at the trigger
input of the analog interface.
In order to activate the output, either
the trigger control signal must be set
to low, or trigger input action must be
set to another function, after which
the OUTPUT ON command must be
re-executed.
Err 31
CMD error (remote)
Err 32
CMD error (remote)
Adhere to calibration sequence
Err 33
Uoff, Ufs
➭
Device must be in voltage
regulating mode
e.g. eliminate short-circuit or reduce
load
Err 34
Iff, Ifs
➭
Device must be in current
regulating mode
e.g. increase load or short circuit the
output terminals
Err 35
Inadequate balancing range
– Check entered data
– Hardware error?
Err 36
Calibration error:
An uncorrected error has occurred, no
CAL data could be saved.
Additional message(s):
40 U_offset
setpoint
41
U_offset measured value
42 U_full-scale
setpoint
43
U_full-scale measured value
44 I_offset
setpoint
45
I_offset measured value
46 I_full-scale
setpoint
47
I_full-scale measured value
Err 99
POWERFAIL
Voltage dip has occurred at auxiliary
supply power
➭
RE-START
+OL or
–OL
Display indicates measuring function
over-ranging:
If measured output voltage UOUT or
output current IOUT violates the
specified range (
➭
+OL or –OL appears at the display.
The appearance of this message
always indicates that the specified
limit values for output voltage or
current have been violated.
Examples:
UOUT indOL:
e.g. caused by
overdriving output voltage with an
analog control signal applied to the
analog interface.
UOUT indicates –OL:
due to
incorrectly connected sensing leads.
IOUT indOL:
e.g. caused by
overdriving output current with an
analog control signal applied to the
analog interface.
POUT indOL:
Since POUT is
calculated by multiplying UOUT and
IOUT, it can be assumed that one of
the above listed causes is responsible
for this message.
Code
Meaning / Cause
Remedy