PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
5
☞
SET FILTER Time
SET FAIL LEVEL Mode
(off, 4 mA or 20 mA)
. Use
arrow keys
to set the value
at which 20 mA is desired, and press
ENTER key
to
enter it.
NOTE:
If the same values are set for 4 mA and 20 mA, the
output automatically goes to and remains at 20 mA.
A time constant (in seconds) can be set to filter or “smooth
out” the analog output signal. A minimum value of “0 sec-
onds” has no smoothing effect. A maximum value of “60
seconds” provides maximum smoothing. Deciding what out-
put filter time to use is a compromise. The higher the filter
time, the longer the analog output signal response time will
be to a change in the measured value.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust to a
desired filter time, and press
ENTER key
to enter it.
When a “WARNING CHECK STATUS” message appears,
indicating that a system problem may exist, the analog out-
put can be set to respond in one of three ways:
•
OFF:
Output remains active.
•
4mA:
Output automatically goes to and remains at 4 mA.
•
20mA:
Output automatically goes to and remains at 20 mA.
To SET FAIL LEVEL mode to suit your application:
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
. Use
×
×
and
Ø
Ø
keys
to select a response mode (OFF, 4mA or
20mA), and press
ENTER key
to enter it.