PART THREE - OPERATION
SECTION 4 - TRANSMITTER CALIBRATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
5
☞
TABLE E -- Conductivity Reference Solutions
Desired Solution Value
Grams NaCl
μS/cm
mS/cm
ppm (NaCl)*
To Be Added
200
500
1000
0.20
0.50
1.00
100
250
500
0.10
0.25
0.50
2000
3000
4000
2.00
3.00
4.00
1010
1530
2060
1.01
1.53
2.06
5000
8000
10,000
5.00
8.00
10.00
2610
4340
5560
2.61
4.34
5.56
20,000
50,000
100,000
20.00
50.00
100.00
11,590
31,950
72,710
11.59
31.95
72.71
*When using ppm measuring scale for compounds other than
NaCl, refer to appropriate chemistry handbook for reference
solution formulation.
2.
Thoroughly rinse the clean sensor in de-ionized water.
Then immerse the sensor in the prepared reference
solution.
Important: Allow the sensor and solution
temperatures to equalize.
Depending on their tem-
perature differences, this may take up to 30 minutes.
NOTE:
Suspend the sensor to prevent it from touching
the container.
Simply laying it into the con-
tainer will produce calibration error.
If the
sensor is tee-mounted, use a smaller container.
Ideally, convert a tee of the same size and
material as the mounting tee into a calibration
container by sealing two of its ends.
3.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
4.
Press
ENTER key
to display
.
5.
Press
ENTER key
again to display
.
6.
Press
ENTER key
again to display a screen like
.
The default 25
°
C reference
temperature is suitable for most applications.
For
another reference, use
arrow keys
to adjust to a differ-
ent temperature. In either case, press the
ENTER key
.