PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
4
Select
TEMP ELEMENT
Type
☞
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.
Create the desired notation on the second line:
A.
Starting with extreme left character position, use
×
×
and
Ø
Ø
keys
to select the desired first character.
B.
Press
Ö
Ö
key
once to select the next character, and
use
×
×
and
Ø
Ø
keys
to select its desired character.
C. Repeat procedure until desired notation is displayed.
3.
Press
ENTER key
to enter the displayed notation.
The temperature element type is factory-set to “PT1000” for
automatic temperature compensation (defines built-in tem-
perature element in GLI electrodeless conductivity sensors).
NOTE:
When “PT1000” is selected but the element is not
connected to the transmitter, a “WARNING:
CHECK STATUS” message will appear. To prevent
or clear this message, connect the element or se-
lect “MANUAL.”
To configure the transmitter for fixed MANUAL temperature
compensation you must select “MANUAL” and enter a spe-
cific temperature:
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display
.