PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
4
SET T FACTOR
(sensor’s GLI-certified
“T” factor)
3.
Press
ENTER key
again to display
.
Use
Ø
Ø
key
to select “MANUAL” for fixed manual temper-
ature compensation, and press
ENTER key
to enter it:
4.
Now determine and enter a specific manual tempera-
ture compensation value:
A.
With the
screen displayed,
press
Ø
Ø
key
once to display
.
B.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust to
a desired temperature for fixed MANUAL compen-
sation, and press
ENTER key
to enter it.
GLI tests each sensor to provide a unique, certified tem-
perature T FACTOR because:
•
Temperature greatly affects conductivity measurement
accuracy.
•
The inherent ohm value of the Pt 1000 RTD temperature
element varies slightly from sensor to sensor, affecting
temperature measurement accuracy.
By entering the sensor’s unique T FACTOR, the transmitter
will provide the highest possible measuring accuracy for
both temperature and conductivity.
1.
With the
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the
displayed value to exactly match the sensor’s GLI-
certified T FACTOR, and press
ENTER key
to enter it.