PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
4
☞
SET FILTER Time
ENTER NOTE (top line
of MEASURE screen)
F.
After all X and Y coordinate values are entered for
each data point in the table, press
ESC key
once to
display
.
G. Press
ENTER key
to display
.
H. Press
ENTER key
again to save the table.
NOTE:
If the table contains unacceptable coordinate
values, the display shows a “CONFIRM FAIL-
URE” message. Pressing
ENTER key
displays
the unacceptable coordinate(s).
A time constant (in seconds) can be set to filter or “smooth
out” the sensor signal. A minimum value of “0 seconds” has
no smoothing effect. A maximum value of “60 seconds” pro-
vides maximum smoothing. Deciding what sensor signal
filter time to use is a compromise. The higher the filter time,
the longer the sensor signal response time will be to a
change in the actual process value.
1.
With the
or
screen displayed, press
Ø
Ø
key
once to display
.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust to a
desired filter time, and press
ENTER key
to enter it.
The top line of the MEASURE screen readouts that sepa-
rately show the measurement, temperature, and analog
output values are factory set to read “COND.” This notation
can be changed, for example, to “BASIN 1” to tailor the
transmitter MEASURE screen to the application. The top
line would then be “MEASURE BASIN 1.” The notation is
limited to eight characters which can be a combination of
capital letters A through Z, numbers 0 through 9, spaces,
# symbols, hyphens, and periods.