PART THREE - OPERATION
SECTION 4 - TRANSMITTER CALIBRATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
6
☞
☞
B. Thoroughly rinse the clean sensor in de-
ionized water. Then immerse the sensor in the
prepared reference solution.
Important: Allow
the sensor and solution temperatures to
equalize.
Depending on their temperature dif-
ferences, this may take up to 30 minutes.
NOTE:
Suspend the sensor to prevent it from
touching the container.
Simply laying
it into the container will produce
calibration error.
If the sensor is tee-
mounted, use a smaller container. Ide-
ally, convert a tee of the same size and
material as the mounting tee into a
calibration container by sealing two of
its ends.
2.
Press
MENU key
to display a “MAIN MENU” screen.
If the
screen is not showing, use
Ø
Ø
or
×
×
key
to display it.
3.
Press
ENTER key
to display
.
4.
Press
ENTER key
again to display
.
5.
Press
ENTER key
to display
.
With the sensor in the process (or % concentration ref-
erence solution), press
ENTER key
again to confirm.
This active
screen appears show-
ing the measurement reading.
NOTE:
During calibration, the analog output is auto-
matically “held” at the last measured value.
6.
Wait for the reading to stabilize which may take up to
30 minutes. Then press
ENTER key
. The “PLEASE
WAIT” screen may appear if the reading is still too un-
stable. After the reading has stabilized, this static
screen appears showing the “last
measured” value.