PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
PRO-series Model E3 Electrodeless Conductivity Transmitter
21
S
ECTION
3
☞
☞
Figure 2-4 shows the terminal block arrangement and
terminal designations for the transmitter.
NOTE:
All terminals are suitable for single wires up to 14
AWG (2.5 mm
2
).
Wiring Tip!
To comply with European Community (CE) electromag-
netic compatibility requirements, follow these general wiring guidelines:
1. Locate transmitter as far as possible from motors and other non-
CE certified devices with excessive electromagnetic emissions.
2. Use GLI-specified ferrites and cables. Failure to do so may elimi-
nate compliance.
Locate all ferrites as close as possible to the
transmitter.
◆
DC Power Supply Cable (GLI 1W0980 two-conductor plus
shield): Connect cable shield to earth ground at the supply
end. Loop cable 2-1/2 times through ferrite (Steward
#28B0686-200, Fair-Rite Corp. #2643665702 or equivalent).
◆
Analog mA Output Cable (four-wire hookup only -- GLI
1W0980 two-conductor plus shield): Connect cable shield to
earth ground at the supply end. Loop cable 2-1/2 times
through ferrite (Steward #28B0686-200, Fair-Rite Corp.
#2643665702, or equivalent).
TB1
TB2
PRO-E-ESP: 0E Conductivity
Shield
S
Red
(Temp)
Yellow
(Signal)
White
(Drive)
(S
ee m
a
n
u
a
l f
o
r 3,
4 W
ir
e
C
onnec
ti
o
n
s
)
Se
rial
#
: EG
0
123
456
2-W
ire C
onnec
tion
3
(-)
6
7
5
4
14-30 V
D
C
C
las
s 2 20mA
max
(+)
1
2
3
6
NC
Green
7
5
4
1
2
Blue
(Drive)
(Ground)
FIGURE 2-4 Transmitter Terminal Designations
ELECTRICAL CONNECTIONS