PART TWO - INSTALLATION
SECTION 1 - UNPACKING
PRO-series Model E3 Electrodeless Conductivity Transmitter
17
P A R T T W O - I N S T A L L A T I O N
S
ECTION
1
Unpack and examine the equipment even if you do not use
it immediately. If there is evidence of damage, notify the
transit carrier immediately.
Recommendation: Save the
shipping carton and packing materials in case the in-
strument must be stored or re-shipped.
S
ECTION
2
2.1 Location
☞
1.
It is recommended to locate the transmitter as close as
possible to the installed sensor. The maximum allow-
able distance between an installed sensor and the
transmitter depends upon the full-scale value you set
for the transmitter measuring range:
200-2000 μS/cm
Full-scale Value
2000-2,000,000 μS/cm
Full-scale Value
200 feet (61 m) max.
300 feet (91 m) max.
NOTE:
When measuring % concentration, convert the
transmitter full-scale value to conductivity to
determine the maximum distance.
The transmitter can be located in a Class 1,
Div.2 hazardous area.
2.
Mount the transmitter in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-4 to +140
°
F or
-20 to +60
°
C).
CAUTION:
EXPOSING THE TRANSMITTER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED LIMIT,
AND DECREASE DISPLAY VISIBILITY.
UNPACKING
MECHANICAL REQUIREMENTS