PART THREE - OPERATION
SECTION 3 - TRANSMITTER CONFIGURATION
PRO-series Model E3 Electrodeless Conductivity Transmitter
5
TABLE D -- Transmitter Configuration Settings (Ranges/Choices and Defaults) -- continued
Displayed Screen Title
Entry Range or Choices (where applicable)
Factory Default
Your Setting
SENSOR Settings (continued)
TEMP ELEMENT:
SET T FACTOR?
950-1050 ohms
1000 ohms
TEMP ELEMENT:
SET MANUAL?
0.0-200.0
°
C
25.0
°
C
TEMPERATURE Display Setting
SET
°
C OR
°
F?
°
C or
°
F
°
C
OUTPUT Settings
SET PARAMETER?
SENSOR or TEMPERATURE
SENSOR
SET 4mA VALUE?
CONDUCTIVITY:
μS/cm: 0-200.0, or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000
S/cm:
0-2.000
CONCENTRATION: 0-99.99% or 0-200.0%
TDS: 0-9999 ppm
TEMPERATURE:
-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
μS/cm: 0
mS/cm: 0
S/cm:
0
CONC: 0.00% or 0.0%
TDS: 0 ppm
TEMPERATURE:
0.0
°
C or 32.0
°
F
SET 20mA VALUE?
CONDUCTIVITY:
μS/cm: 0-200.0 or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000
S/cm:
0-2.000
CONCENTRATION: 0-99.99% or 0-200.0%
TDS: 0-9999 ppm
TEMPERATURE:
-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
CONDUCTIVITY:
μS/cm: 200.0 or 2000
mS/cm: 2.000, 20.00, 200.0
or 2000
S/cm:
2.000
CONC: 99.99% or 200.0%
TDS: 9999 ppm
TEMPERATURE:
100.0
°
C or 212.0
°
F
SET FILTER?
0-60 seconds
0 seconds
SET FAIL LEVEL?
OFF, 4 mA or 20 mA
OFF
PASSCODE Setting
SET PASSCODE?
DISABLED or ENABLED
DISABLED
TEST/MAINT Simulation Function Settings
SELECT SIM?
SENSOR or TEMPERATURE
SENSOR
SIM SENSOR?
CONDUCTIVITY:
μS/cm: 0-200.0, or 0-2000
mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000
S/cm:
0-2.000
CONCENTRATION: 0-99.99% or 0-200.0%
TDS: 0-9999 ppm
TEMPERATURE:
-20.0 to +200.0
°
C or -4.0 to 392.0
°
F
Present measured value of
selected parameter