PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
56
ACTIVATION
(configuration values)
☞
The group of configuration settings available to a relay is
dependent on its selected function mode (ALARM or
CONTROL).
Relays set for STATUS function are not
configurable.
Table B describes all relay configuration set-
tings, categorized by relay function mode:
Table B -- RELAY CONFIGURATION SETTINGS
Setting
Description
For ALARM Relay
Low Alarm
Sets the value at which the relay will turn on in
response to decreasing measured value.
High Alarm
Sets the value at which the relay will turn on in
response to increasing measured value.
Low
Deadband
Sets the range in which the relay remains on after the
measured value increases above the low alarm value.
High
Deadband
Sets the range in which the relay remains on after the
measured value decreases below the high alarm value.
Off Delay
Sets a time (0-300 seconds) to delay the relay from
normally turning off.
On Delay
Sets a time (0-300 seconds) to delay the relay from
normally turning on.
For CONTROL Relay
Phase
A “high” phase assigns the relay setpoint to respond
to increasing measured value; conversely, a “low”
phase assigns the relay setpoint to respond to de-
creasing measured value.
Setpoint
Sets the value at which the relay will turn on.
Deadband
Sets the range in which the relay remains on after the
measured value decreases below the setpoint value
(high phase relay) or increases above the setpoint
value (low phase relay).
Overfeed
Timer
Sets the time (0-999.9 min.) to limit how long the re-
lay can remain “on.” For more details on overfeed
timer operation, see Part Three, Section 7.
Off Delay
Sets a time (0-300 seconds) to delay the relay from
normally turning off.
On Delay
Sets a time (0-300 seconds) to delay the relay from
normally turning on.
For STATUS Relay
No settings available -- status relay cannot be configured.
NOTE:
It is possible to enter values that always keep a re-
lay active or inactive. To avoid this, be sure that
“low” values are lower than “high” values.
When a relay is set for STATUS function, the
symbol at the start of the “ACTIVATION” line de-
notes that this menu item is not available.