PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
46
1.
With the
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select the “TEMP ELEMENT” line.
2.
Press
ENTER key
to display
6(7 7 )$&725
6(7 0$18$/
(;,7
.
3.
With the “SELECT TYPE” line selected, press
ENTER
key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view the three choices:
•
PT1000:
Configures analyzer for use with a Pt 1000
RTD temperature element (used in all GLI Model
3400-series contacting conductivity sensors).
•
PT100:
Configures analyzer for use with a Pt 100
RTD temperature element.
•
MANUAL:
Configures analyzer for fixed manual
temperature compensation when not using a
temperature element.
4.
With the desired choice displayed, press
ENTER key
to
enter this selection. When “MANUAL” was selected,
you must set the specific manual temperature compen-
sation value:
A.
With the
6(7 7 )$&725
6(7 0$18$/
(;,7
screen displayed, use
Ø
Ø
key
to select the “SET MANUAL” line.
B.
Press
ENTER key
to display a screen like
. Use
arrow keys
to adjust the dis-
played value to the desired fixed temperature, and
press
ENTER key
to enter it.