PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
51
SET 0/4 mA and
20 mA VALUES
(range expand)
☞
1.
With the
6(7 287387
6(7 5(/$< $
6(7 5(/$< %
6(7 5(/$< &
6(7 5(/$< '
6(7 3$66&2'(
6(7 & 25 )
/$1*8$*(
6(1625 $
6(1625 %
&$/&8/$7,21
(;,7
screen displayed and the
“SET OUTPUT 1” line selected, press
ENTER key
to
display
6(7 P$ 9$/8(
6(7 P$ 9$/8(
6(7 75$16)(5
6(7 ),/7(5
6&$/( P$P$
(;,7
.
2.
With the “SET PARAMETER” line selected, press
ENTER key
to display a screen like
.
Use
Ø
Ø
and
×
×
keys
to view the choices. With the de-
sired choice displayed, press
ENTER key
to enter this
selection.
You can set parameter (or calculated measurement) values
to define the endpoints at which the minimum and maximum
output values are desired.
1.
With the
6(7 P$ 9$/8(
6(7 P$ 9$/8(
6(7 75$16)(5
6(7 ),/7(5
6&$/( P$P$
(;,7
screen displayed, use
Ø
Ø
key
to select the “SET 4 mA VALUE” line.
2.
Press
ENTER key
to display a screen like
. Use
arrow keys
to set the displayed
value at which 0/4 mA is desired, and press
ENTER
key
to enter the value.
3.
After the
6(7 P$ 9$/8(
6(7 P$ 9$/8(
6(7 75$16)(5
6(7 ),/7(5
6&$/( P$P$
(;,7
screen re-appears, use
Ø
Ø
key
to select the “SET 20 mA VALUE” line.
4.
Press
ENTER key
to display a screen like
. Use
arrow keys
to set the displayed
value at which 20 mA is desired, and press
ENTER key
to enter the value.
NOTE:
If the same values are set for 0/4 mA and 20 mA, the
output automatically goes to, and remains at, 20 mA.