PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Rev. 5-1000
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
45
ENTER NOTE (top line
of MEASURE Screen)
Select
TEMP ELEMENT
Type
☞
The top line of the MEASURE screen readouts that sepa-
rately show the measurement are factory set to read “COND
A” and “COND B.” These notations can be changed, for ex-
ample, to “BASIN 1” and “BASIN 2” to tailor the analyzer
MEASURE screen to the application. The respective top
lines would then be “MEASURE BASIN 1” and “MEASURE
BASIN 2.” The notation is limited to eight characters which
can be a combination of capital letters A through Z, num-
bers 0 through 9, and spaces.
1.
With the
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select the “ENTER NOTE” line.
2.
Press
ENTER key
to display
. Create
the desired notation on the second line:
A.
Starting with extreme left character position, use
×
×
and
Ø
Ø
keys
to select the desired first character.
B.
Press
Ö
Ö
key
once to select the next character, and
use
×
×
and
Ø
Ø
keys
to select its desired character.
C. Repeat this procedure until the desired notation is
displayed.
3.
Press
ENTER key
to enter the displayed notation.
Configure the analyzer for either automatic temperature
compensation (uses Pt 1000 ohm RTD built into sensor), or
fixed MANUAL temperature compensation. When using
MANUAL you must determine and enter a specific tem-
perature value.
NOTE:
When a temperature element type has been se-
lected but the element is not connected to the
analyzer, a “WARNING CHECK STATUS” message
will appear. To prevent or clear this message, con-
nect the element or select “MANUAL.”