PART THREE - OPERATION
SECTION 5 - ANALYZER CALIBRATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
70
5.5 Analog Outputs
(1 and 2) Calibration
☞
The analyzer analog outputs are factory-calibrated. How-
ever, they can be re-calibrated at any time if desired.
These
instructions calibrate Output 1. Calibrate Output 2 in the
same way using its respective menu screens.
NOTE:
When the passcode feature is enabled (Section
4.7), you must successfully enter the passcode
before attempting to calibrate the analog outputs.
When an output is configured to be 0-20 mA, the
analyzer will calibrate the 4 mA and 20 mA values
(not the 0 mA value). Also, the analyzer adjustment
range for output values during calibration is ± 2 mA.
1.
Press
MENU key
to display
&21),*85(
7(670$,17
(;,7
.
2.
With the “CALIBRATE” line selected, press
ENTER key
to display
6(1625 %
2873876
(;,7
.
3.
Use
Ø
Ø
key
to select the “OUTPUTS” line, and press
ENTER key
to display
.
4.
With the “CAL OUTPUT 1” line selected, press
ENTER key
to display
.
5.
With the “CAL OUT 1 4 mA” line selected, press
ENTER key
to display a screen like
.
The displayed value is “counts” -- not mA -- that dy-
namically change when the output is adjusted.
6.
Use a calibrated digital multimeter to measure Output
1’s actual minimum value provided at Terminals 2 and 3
on TB1.