PART THREE - OPERATION
SECTION 4 - ANALYZER CONFIGURATION
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
40
Select Temperature
COMPENSATION
☞
3.
With the desired choice displayed, press
ENTER key
to
enter this selection.
Configure the required type of temperature compensation
for the selected measurement.
1.
With the
',63/$< )250$7
&203(16$7,21
&21),* &21'8&7
&21),* /,1($5
&(// &2167$17
6(7 ),/7(5
38/6( 68335(66
(17(5 127(
7(03 (/(0(17
(;,7
screen displayed, use
Ø
Ø
key
to select “COMPENSATION” line.
2.
Press
ENTER key
to display a screen like
. Use
Ø
Ø
and
×
×
keys
to view choices:
•
LINEAR (recommended for most aqueous solutions)
•
AMMONIA (not shown for TDS measurement; only
use this built-in temperature properties table for
specific applications -- consult factory)
•
NATURAL WATER (not shown for TDS measure-
ment; only use this built-in temperature properties
table for specific applications -- consult factory)
•
OPTIONAL TABLE* (user-specified temperature table)
•
NONE (measured values are uncompensated)
* Only select OPTIONAL TABLE choice when the analyzer is
equipped with a customer-specified, factory-configured tem-
perature compensation table, or you intend to configure a
special table using the optional GLI software tool kit
1000G3311 and an IBM-compatible PC.
NOTE:
The factory default for temperature compensa-
tion is LINEAR with a 2.00% per
°
C slope and
25.0
°
C reference temperature. This provides
the best results for most aqueous solutions. To
enter different slope and reference temperature
values for an uncommon solution, refer to
subheading “CONFIG LINEAR Temperature
Compensation” for details.
3.
With the desired choice displayed, press
ENTER key
to
enter this selection.