PART TWO - INSTALLATION
SECTION 3 - ELECTRICAL CONNECTIONS
Model C53 Conductivity Analyzer (universal-mount 1/2 DIN)
Rev. 5-1000
26
☞
The analyzer can be used with one or two conductivity sen-
sors. Refer to Figure 2-5 or 2-6 and connect the wires of
each sensor (or interconnect) cable to appropriate terminals
on TB1, matching colors as indicated.
NOTE:
For best immunity to electromagnetic interference,
always connect each sensor cable’s outer shield
wire (clear with black band -- not its clear-only inner
shield wire) to:
•
Grounding strip at bottom of case (5 open holes)
for analyzers with “B” prefix serial number.
•
Terminal 17 and Terminal 11 on TB1 for each
respective sensor for analyzers with “A” or no
letter prefix serial number.