Publication 31231, Rev. J, October 2003
4-1
4
Performance Verification
4.1
Introduction
•
This chapter provides step-by-step procedures to verify performance of the Series 12000A
Microwave Synthesizers for all three models (Modulation related items are in Section 4.4 for the
Series 125XXA/127XXA models)
•
Sectional header page references are only provided for items pertaining directly to performance
verification testing
•
The required warm-up time before testing is 72 hours. The warm-up period can be reduced to
30 minutes if timebase accuracy is not to be tested for all three models
•
In these procedures, the instrument being tested may be referred to as the 12000A or UUT (Unit
Under Test) for all three models
4.1.1
Recommended Equipment
The following equipment is recommended before starting the performance test routines (the
recommended equipment can be substituted provided the specifications are sufficiently compatible):
Introduction
(Performance Verification)
Model
124XXA
125XXA/
127XXA
4.1.1 - Recommended Equipment
√
√
Table 4-1: Series 12000A Recommended Equipment
Equipment
Example
Frequency Standard
10 MHz Stanford Research FS 725
Oscilloscope
Tektronix TDS3052B or Equivalent
Microwave Frequency Counter
XL Microwave 3260 or Equivalent
Power Meter and Sensor
Giga-tronics 8541C/Series 8650A w/ 80313A Sensor or Equivalent
Spectrum Analyzer
HP 8566B or Equivalent
Detector
Herotek Model DZ262-44 or Equivalent
Function (Audio) Generator
Stanford Research DS-345 or Equivalent
Measuring Receiver
HP 8902A or Equivalent
Distortion Analyzer
HP 339A or Equivalent
L.O. Generator
Giga-tronics Microwave Synthesizer 12720A or Equivalent
Universal Counter
Racal Dana 2201 or Equivalent
Mixer IF Amplifier w/ Divide-by-40 Output
Giga-tronics Mixer/Divider (P/N 002CA04900)
Mixer (For Attenuator Test)
Marki M40020MJ
Filter 21.4 MHz
Mini-Circuits SVP 21.4
Fixed Attenuator 3 dB
Midwest ATT-0263-03-SMA-02
Fixed Attenuator 10 dB
Midwest ATT-0263-10-SMA-02
Summary of Contents for Series 12000A
Page 3: ...DECLARATION OF CONFORMITY ...
Page 4: ......
Page 11: ...Preface Publication 31231 Rev J October 2003 vii ...
Page 14: ...Series 12000A Microwave Synthesizers x Publication 31231 Rev J October 2003 ...
Page 18: ...Series 12000A Microwave Synthesizers xiv Publication 31231 Rev J October 2003 ...
Page 20: ...Series 12000A Microwave Synthesizers xvi Publication 31231 Rev J October 2003 ...
Page 22: ...Series 12000A Microwave Synthesizers xviii Publication 31231 Rev J October 2003 ...
Page 24: ...Series 12000A Microwave Synthesizers xx Publication 31231 Rev J October 2003 ...
Page 62: ...Series 12000A Microwave Synthesizers 1 38 Publication 31231 Rev J October 2003 ...
Page 234: ...Series 12000A Microwave Synthesizers B 2 Publication 31231 Rev J October 2003 ...