A-1
GXDP-300
Appendix A. Technical Information
Appendix A. Technical Information
A.1 Technical specifi cations
Type
GXDP-300
Classifi cation
Complies with IEC 60601-1/1995, IEC 60601-2-7/1998, IEC 60601-2-28/1993 and
IEC 60601-2-32/1994, IEC 878
UL 2601-11/2006 (for products with the UL Classifi cation Mark) and EN 55011
standards
Conforms with the regulations of DHHS Radiation Performance Standard, 21CFR
Subchapter J.
Safety according to IEC 60601-1
Protection against electric shock - Class 1
Degree of protection - Type B applied with no conductive connection to the patient
Protection against the ingress of liquids - IPX 0
Disinfection methods:
- mild soapy water (non-abrasive)
- non-alcohol based disinfectant for the chin rest
- disposable plastic covers for bite guide, chin rest and lip support
For use in environments where no fl ammable anaesthetics nor fl ammable cleaning
agents are present
Mode of operation - continuous operation/intermittent loading
Unit description
A dental panoramic x-ray units with a high frequency switching mode x-ray
generator. The unit takes panoramic exposures.
The unit uses a CCD sensor as an image receptor.
Generator
TUBE
Toshiba D-052 SB or D-054 SB or CEI 105/5 or equavalent
TUBEHEAD HOUSING ASSEMBLY
THA-M-3
FOCAL SPOT
0.5 mm (IEC 60336/1993)
FOCAL SPOT ACCURACY
The accuracy is 10mm from the marking on the tubehead cover
TARGET ANGLE
5º
A-1
GXDP-300
Appendix A. Technical Information
Appendix A. Technical Information
A.1 Technical specifi cations
Type
GXDP-300
Classifi cation
Complies with IEC 60601-1/1995, IEC 60601-2-7/1998, IEC 60601-2-28/1993 and
IEC 60601-2-32/1994, IEC 878
UL 2601-11/2006 (for products with the UL Classifi cation Mark) and EN 55011
standards
Conforms with the regulations of DHHS Radiation Performance Standard, 21CFR
Subchapter J.
Safety according to IEC 60601-1
Protection against electric shock - Class 1
Degree of protection - Type B applied with no conductive connection to the patient
Protection against the ingress of liquids - IPX 0
Disinfection methods:
- mild soapy water (non-abrasive)
- non-alcohol based disinfectant for the chin rest
- disposable plastic covers for bite guide, chin rest and lip support
For use in environments where no fl ammable anaesthetics nor fl ammable cleaning
agents are present
Mode of operation - continuous operation/intermittent loading
Unit description
A dental panoramic x-ray units with a high frequency switching mode x-ray
generator. The unit takes panoramic exposures.
The unit uses a CCD sensor as an image receptor.
Generator
TUBE
Toshiba D-052 SB or D-054 SB or CEI 105/5 or equavalent
TUBEHEAD HOUSING ASSEMBLY
THA-M-3
FOCAL SPOT
0.5 mm (IEC 60336/1993)
FOCAL SPOT ACCURACY
The accuracy is 10mm from the marking on the tubehead cover
TARGET ANGLE
5º
Reviewed: Westling Lauri 22.06.11 11:27:45
Approved: Weckman Janne 22.06.11 13:46:25
See the PaloDEx Group Oy PDM system to determine the status of this document. Printed out: 05.09.11 15:25:27
D506222
Approved