Appendix A. Technical Information
System requirements and connections
- The PC and any other external device(s) connected to the system must meet
the IEC 60950 standard (minimum requirements). Devices that do not meet the
IEC 60950 standard must not be connected to the system as they may pose a
threat to operational safety.
- The PC and any other external devices must be connected in accordance with
IEC 60601-1-1.
- The x-ray unit must be connected to it’s own separate power supply. The PC
and any other external devices must NOT be connected to the same power
supply as the x-ray unit.
- Position the PC and any other external device at least 1.85 m (73”) from the
x-ray unit so that the patient cannot touch the PC or any other external device
while being x-rayed.
- The PC and any other external devices shall not be connected to an extension
cable.
- Multiple extension cables shall not be used.
- Do not position the PC where it could be splashed with liquids.
- Clean the PC in accordance with the manufacturer’s instructions.
X-ray system - to IEC 60601-1-1
A-5
GXDP-300
Appendix A. Technical Information
System requirements and connections
- The PC and any other external device(s) connected to the system must meet
the IEC 60950 standard (minimum requirements). Devices that do not meet the
IEC 60950 standard must not be connected to the system as they may pose a
threat to operational safety.
- The PC and any other external devices must be connected in accordance with
IEC 60601-1-1.
- The x-ray unit must be connected to it’s own separate power supply. The PC
and any other external devices must NOT be connected to the same power
supply as the x-ray unit.
- Position the PC and any other external device at least 1.85 m (73”) from the
x-ray unit so that the patient cannot touch the PC or any other external device
while being x-rayed.
- The PC and any other external devices shall not be connected to an extension
cable.
- Multiple extension cables shall not be used.
- Do not position the PC where it could be splashed with liquids.
- Clean the PC in accordance with the manufacturer’s instructions.
X-ray system - to IEC 60601-1-1