User Manual
23
GXDP-300
®
5. Troubleshooting and Maintenance
Error symbols
REASON
i.
The PC connected to the unit is not on.
ii.
GxPicture in the PC is not open.
iii.
The cable connecting the unit to the PC is disconnected or damaged.
iv.
The IP address is not set correctly.
SOLUTION
i.
Switch the PC on.
ii. Open
GxStart.
iii.
Reconnect the cable. If damaged, contact service.
iv. Reconfi gure the IP address.
REASON
The emergency stop button is pressed down in the STOP position.
SOLUTION
Rotate and release the emergency stop button. The error symbol will clear.
User errors
H1
REASON
The exposure button was released during an exposure.
SOLUTION
Clear the error message and check if the attempted exposure is suffi cient for the
diagnostic task. If it is not, take a new exposure.
If the exposure failed while the exposure button was still being pressed, check the
exposure switch by taking a test exposure without patient to see if the exposure
button is defective or not. If the same problem occurs again, contact service.
User Manual
23
GXDP-300
®
5. Troubleshooting and Maintenance
Error symbols
REASON
i.
The PC connected to the unit is not on.
ii.
GxPicture in the PC is not open.
iii.
The cable connecting the unit to the PC is disconnected or damaged.
iv.
The IP address is not set correctly.
SOLUTION
i.
Switch the PC on.
ii. Open
GxStart.
iii.
Reconnect the cable. If damaged, contact service.
iv. Reconfi gure the IP address.
REASON
The emergency stop button is pressed down in the STOP position.
SOLUTION
Rotate and release the emergency stop button. The error symbol will clear.
User errors
H1
REASON
The exposure button was released during an exposure.
SOLUTION
Clear the error message and check if the attempted exposure is suffi cient for the
diagnostic task. If it is not, take a new exposure.
If the exposure failed while the exposure button was still being pressed, check the
exposure switch by taking a test exposure without patient to see if the exposure
button is defective or not. If the same problem occurs again, contact service.