EQUIPMENT OVERVIEW: Technical Specifications
Revision J
Solar 7000/8000/View Patient Monitor
2-15
414993-001
Environmental Specifications
Item
Description
Power Requirements
Solar 7000 Color Monitor:
■
110 ± 11 VAC, 50/60-Hz, single phase
■
120 ± 12 VAC, 50/60-Hz, single phase
■
220-230 ± 22 VAC, 50/60-Hz, single phase
■
240 ± 24 VAC, 50/60-Hz, single phase
Solar 7000 Monochrome: Monitor and Solar 8000 Processing Unit
■
110 ± 20 VAC, 50/60-Hz, single phase
■
220-230 ± 40 VAC, 50/60-Hz, single phase
Power Consumption
Solar 7000 Color Monitor: 200 W (180 W for a Tram-rac housing with power
supply connected)
Solar 7000 Monochrome Monitor: 120 W (100 W for a Tram-rac housing with
power supply connected)
Solar 8000 Processing Unit: 100W (maximum)
SolarView Remote Display Controller: 25W (maximum)
Low Voltage Shutdown
Solar 7000 Color Monitor: 88 VAC/106 VAC/196 VAC/214 VAC
Solar 7000 Monochrome Monitor: 85 VAC/170 VAC
Solar 8000 Processing Unit: 90 VAC/190 VAC
SolarView Remote Display Controller: 90 VAC/190 VAC
Cooling
Solar 7000 Color Monitor: Forced convection
Solar 7000 Monochrome Monitor and Solar 8000 Processing Unit: Natural
convection
Heat Dissipation
Solar 7000 Color Monitor: 680 Btu/hr (200 W)
Solar 7000 Monochrome Monitor: 409 Btu/hr (120 W)
Solar 8000 Processing Unit: 100 Btu/hr (30W)
SolarView Remote Display Controller: 50 Btu/hr (15W)
Operating Conditions
■
Ambient
Temperature
■
Relative Humidity
Solar 7000 Monitor: 10°C to 35°C (50°F to 95°F)
Solar 8000 Processing Unit: 10°C to 40°C (50°F to 104°F)
SolarView Remote Display Controller: 10°C to 40°C (50°F to 104°F)
Solar 7000 Monitor: 15% to 95% (noncondensing)
Solar 8000 Processing Unit: 15% to 95% (noncondensing)
SolarView Remote Display Controller: 15% to 95% (noncondensing)
Storage Conditions
■
Temperature
■
Relative Humidity
Solar 7000 Monitor: –10°C to 50°C (14°F to 122°F)
Solar 8000 Processing Unit: –40°C to 70°C (–40F° to °158F)
SolarView Remote Display Controller: –40°C to 70°C (–40F° to °158F)
Solar 7000 Monitor: 0% to 95% (noncondensing)
Solar 8000 Processing Unit: 15% to 95% (noncondensing)
SolarView Remote Display Controller: 15% to 95% (noncondensing)
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......