MAINTENANCE: Checkout Procedure
3-18
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
3.
Select CROSS HATCH test pattern and check for display linearity.
If the Solar 7000 display is not linear, perform “Display Calibration”
in Chapter 5: “Calibration.”
4.
Press Trim Knob control to exit.
Select VERTICAL GRAY BARS in the service menu. All gray scales
should be visible. The brightest bar should be white and not blurry.
If the Solar 7000 display fails this step, perform “Display
Calibration” in Chapter 5: “Calibration.”
5.
Press Trim Knob control to exit. Select RESTART to exit menu.
For Solar 7000 color monitors only, observe the white color quality of
the display windows in the MAIN MENU. To remove any hue of color,
perform “Display Calibration” in Chapter 5: “Calibration.”
Solar 8000/View Display
Check
For the Solar 8000 or SolarView display, refer to the specific
manufacturer’s documentation.
Tram-rac Housing Check
Do the following to check the Tram-rac housing.
1.
Verify power LED is ON at the Tram-rac housing.
2.
Disconnect and reconnect the Tram-rac housing communication
cable. Verify the recovery of the waveforms.
3.
If the Tram-rac housing has additional slots for input modules, insert
a BP module. Connect simulator and verify communication to the
monitor. Repeat for each slot.
4.
If the Tram-rac housing has an optional power supply, check the
following on the connector that applies to your equipment.
◆
16.5V is
not
present at pin 5 of the TRAM-NET
connector with respect to pin 9.
◆
16.5V is
not
present at pin 5 with respect to chassis
ground of the Tram-rac housing.
5.
The following step does not apply for a Tram-rac 2 housing. Check
the analog output connector (yellow) using an oscilloscope. Observe a
signal at the appropriate pins found in the next table. The output
signal is dependent upon which Tram and input module functions are
activated at the monitor. Tram-rac 3 & 4 housings use the front
round connector.
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......