TROUBLESHOOTING: Service Menus
4-32
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
Copy Unit Defaults
This menu option is used to copy Monitor Defaults from one monitor to
another monitor. Refer to “Transfer Monitor Defaults” in Chapter 6:
“Configuration” for complete details.
Degauss
(Solar 7000 Only)
This menu selection is used to degauss the Solar 7000 display. Select
YES to degauss the color display periodically or whenever the red color
becomes impure (not true red). Note that the display is automatically
degaussed every time the unit is turned ON at the back switch. Once the
display is manually degaussed, it cannot be degaussed again for at least
10 minutes.
Time and Date
This menu selection allows changes to the time and date and may affect
the time and date for the entire monitoring network.
WARNING
This menu should rarely be used because patient
histories will be lost.
This procedure is described in “Programming Time and Date ” in Chapter
6: “Configuration.”
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......