CONFIGURATION: Advanced User Procedures
Revision J
Solar 7000/8000/View Patient Monitor
6-23
414993-001
What Do Error Codes
Mean?
The error log contains more than just operating system errors. Many
events that have an impact upon the system are also entered into the log.
The 700-series of error codes are really system initiated events. Listed
below are some of the event/error codes you might find useful.
NOTE
The monitor may be referred to as a display or scope in
the error code descriptions.
Definition of Parameters
Error Code
Description
400-4FF
Network errors were detected.
703
Diagnostic tests were completed.
70B
Internet address was changed. The network address
for the monitor was changed. This should only be done
by qualified service personnel.
70C
Video test was completed. This test should only be
performed by qualified service personnel.
70D
Rear power switch was turned off.
70E
Time was changed from this monitor. This helps
determine how the system-wide time may have been
altered.
70F
Date was changed from this monitor. This helps
determine how the system-wide date was altered.
710
Incompatible software was detected. If the main
processor software finds that the software operating
on the communication software incompatible, it turns
off the communication (network) controller and enters
this data into the error log. When the monitor won’t
“talk” to the network, looking for this entry in the
error log is one part of the troubleshooting process.
714
Incompatible TRAM module software was detected. If
the main processor software finds the software
revision in the TRAM is incompatible, it turns off
communication with the TRAM and enters an error
log. Refer to the software compatibility listing in the
SOFTWARE REVISION window.
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......