TROUBLESHOOTING: Troubleshooting Software Updates
4-46
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
Troubleshooting Software Updates
Problems and
Solutions
The following is a list of problems commonly encountered during a
software update with their solutions.
Central station contains the software, but the monitor
cannot find it.
The central station has loaded the monitor software onto its hard drive,
but the monitor does not see the central station on the network. There
are two ways to get the central station to broadcast across the network.
•
If patient monitoring may be interrupted, press the CTRL, ALT, and
DELETE keys simultaneously to reboot the central station.
•
If the central station is monitoring patients, do the following:
1.
At the central station, execute the following menu sequence, starting
from the MAIN menu:
CENTRAL SETUP
SERVICE
Enter Password (MEI CS 123)
SERVICE MONITOR
NOTE: The commands presented in italics are case sensitive.
2.
Type ps eaglefs and press the ENTER key. Enabling the EAGLEFS
program that teaches your central station how to broadcast the
software on the network. The central station will respond in one of
two ways:
◆
If the central station is not running EAGLEFS, it will respond
with: INVALID PROCESS ID OR NAME. Type
run eaglefs HDØ
and press the ENTER key.
◆
If the central station lists EAGLEFS as a running process, go
to the next step.
3.
Press the MAIN MENU key on the front panel to exit the SERVICE
MONITOR and go back to the monitor to download the software.
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......