CONFIGURATION: Configuring a Monitor
6-6
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
Programming Care
Unit Name
Up to seven characters are used to identify the care unit. These
characters are displayed at the top right of the screen immediately
preceding the bed number. Do the following procedure to enter a care
unit name.
1.
To access SET UNIT NAME option, execute the following menu
sequence starting from the MAIN menu:
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
MONITOR SETTINGS
UNIT NAME
2.
Note that the first character in the care unit name is highlighted.
Rotate the Trim Knob control to change this character.
3.
When satisfied with the character chosen, press the Trim Knob
control. This causes the highlight to move to the next character.
4.
Follow this procedure for all characters. Up to seven characters may
be entered.
5.
Turn the Trim Knob control to highlight SET UNIT NAME and press
the Trim Knob control to exit.
Programming Bed
Number
The bed number identifies a particular patient bed. Up to five characters
are used to identify bed number. This number will be displayed at the top
right of the screen.
1.
To access SET BED NUMBER option, execute the following menu
sequence starting from the MAIN menu:
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
MONITOR SETTINGS
SET BED NUMBER
2.
Using Trim Knob control, follow same procedure as listed above to
change the bed number.
↓
↑
↓
↑
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......