UPPER LEVEL ASSEMBLY: Solar 8000/View Disassembly Guidelines
7-24
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
Main Processor PCB
Removal
1.
Squeeze and pull the harness clip from the chassis as shown below.
2.
Disconnect J7 harness from the main processor PCB and remove five
screws attaching the PCB.
3.
After removing the PCB, disconnect harness J1 and J8.
4.
Remove the serial EEPROM U9 and insert it into the replacement
main processor PCB. The serial EEPROM must stay with the unit.
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......