CALIBRATION: Solar 7000 Color Monitor
Revision J
Solar 7000/8000/View Patient Monitor
5-7
414993-001
CRT Controller PCB
Calibration
This procedure should be performed whenever a CRT controller PCB has
been repaired in a Solar 7000 color monitor. When this procedure is
completed, perform “Display Calibration” found earlier in this chapter.
The following steps describe a simple method to calibrate the CRT
controller PCB using a special display generated by the monitor circuits.
To perform this procedure, you will need the special tools and items listed
below. Equivalent equipment may be substituted.
WARNING
Beware of AC line voltage at the power module. When
rear power switch is OFF, AC line voltage is still present.
Use an insulated screw driver at all times when making
adjustments.
WARNING
Make sure the monitor is not monitoring a patient,
because monitoring will be interrupted.
Required Tools/Special Equipment
Item
Manufacturer and
Part Number/Model
Very small Blade Screwdriver
–
Oscilloscope (100 MHz BW or greater)
Tektronix 2215
Multimeter (DMM)
Fluke 8060A
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......