CONFIGURATION: Configuring a Monitor
Revision J
Solar 7000/8000/View Patient Monitor
6-5
414993-001
Programming
Patient-Monitor Type
The PATIENT-MONITOR TYPE selection allows the user to select the
type of monitor desired, i.e adult, neonatal or operating room. Different
alarms and parameters are activated for each selection. This menu item
is part of the SERVICE MODE to keep a casual user from changing the
type of the monitor.
CAUTION
Each time the patient-monitor type is changed, the
ADMIT MENU function will default to STANDARD
configuration. Be aware that some alarms and
parameters may be changed.
Do the following procedure to change the PATIENT-MONITOR TYPE.
Refer to the operator’s manual for a full description of alarm, limit, and
parameter changes for each monitor type.
1.
To access PATIENT-MONITOR TYPE option, execute the following
menu sequence starting from the MAIN menu:
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
PATIENT-MONITOR TYPE
2.
Rotate Trim Knob control to select the type of environment the
monitor will be used in.
3.
Press Trim Knob control to exit. Your selection will be displayed at
the top of the screen after the time.
↓
↑
Summary of Contents for Solar 7000
Page 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Page 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Page 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Page 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Page 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Page 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Page 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Page 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Page 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Page 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Page 213: ......
Page 214: ......