Chapter 2 - Power, Ground & Interconnect Cables
Page 127
2 –
Install
Power
8.1.3
HVDC Connection
Note:
Refer to
Table 2-3, System Interconnect Cables on page 100
.
Connect the internally shielded HVDC cable to TS2 on the standing panel. See
for the
location of the connector and
for details. Observe polarities and grounds. Do not cut or
shorten cables unless you have all of the appropriate tools and crimper to re-terminate. If short
cables are needed, have the PMI order the short cable set.
WARNING
Excess cable length cannot be stored under or behind the PDU. If cables are to be stored in the
cable tray, do not overfill. Consult the local electrician to determine the maximum fill rate for your
area.
Figure 2-31 HVDC Connection
Check box when complete.
CIRCUIT
BREAKER
DESCRIPTION
CB2
Circuit Protection (Axial Drive)
CB3
Full Winding Protection
CB4
CT Gantry Service Outlets
CB5
CT Gantry rotating loads
CB6
Table & CT Gantry Stationary Loads
CB7
Operator Console Load
CB9
VMR1, Control P.S Load
CB10
VMR2
Table 2-16 Panel Circuit Breaker Descriptions
Install Shield Clamp tightly
around exposed Cable Shield
at lower left of PDU.
TS2-1
(Red)
TS2-3 (Black)
Ground
+
-
(Shield)
Standing Panel - HVDC SUPPLY
OUTPUT CONNETCTIONS