Chapter 1 - Position Subsystems
Page 79
sys
tems
Figure 1-59 Center tapped holes under mounting holes in table base
11.8
Install the Anchors (For Lite Table)
Recommended - Use “Hilti Kwik-Bolt 3" anchors P/N 2106573-3 (12.7 mm (½") dia. by 140 mm (5-
½") long) as shipped with the system for this procedure.
Figure 1-60 Lite Table Anchor Assembly
1.) Remove the alignment tool and repack the install kit.
2.) Assemble the anchors before you install them. Refer to
.
Bolt centering is important to
provide ± 3mm of adjustment
for electrical alignment.
Always use the drilling
centering tool when drilling all
bolt holes.
100 mm (3.9 in.) or
more
Anchor nut
Table base
Floor
Floor
13mm
(1/2 in.)
41mm (1.6 in.)
or
less
Washer
(P9155BZ)
90 mm (3.5 in.)
Anchor bolt
(2106573-3)