Chapter 1 - Position Subsystems
Page 41
sys
tems
Figure 1-6 Gantry and Table Base Leveling Pads (Starting Positions)
NOTICE
Gantry dollies weigh approximately 250 lbs each. Exercise caution when removing
dollies so as to not damage the floor covering.
6.) Using a 14mm hex socket, remove the dollies from the gantry by removing the three dolly bolts
found at both ends of the gantry (
Figure 1-7 Gantry Dolly Bolts
7.) Remove the dolly plates on both sides of the gantry. Retain the dolly plates in the service
cabinet at the hospital.
Figure 1-8 Dolly Plate