Section 10.0 - Table Installation (Lite Table)
2.) Release a latch by pushing a bracket to move the cradle by hands. See
.
Figure 1-43 Release the Cradle
3.) Center the cradle on the drive rollers assembly by pushing the cradle into the gantry to its
maximum position and back to just before the latch position six times. The cradle should be
centered.
You must push and pull a center of cradle hand. See
.
Figure 1-44 Push and Pull Cradle
10.3.3
Install the Table Cradle Laser Alignment Plates
Required tool: TG-alignment tool 5180876
1.) Install the front table laser alignment which has a cross slot to the front of the cradle.
a.) Loose adjustment knob to make the slot plate free.
b.) The slot in the front of the tool should be over the cradle center-line. See
.
Figure 1-45 Front Table Laser Alignment
Original Position
Release Position
Release Lever