Appendix D - Remote Monitor Option Console Wiring
Page 201
–
2.) With the monitor and computer switched off, remove the video cable from the back of the image
monitor. See
. (The monitor shown below may appear different from the one you
have.) Remove this cable from the HP computer video connector. Discard this cable. If the
monitor has BNC connectors, remove them.
Figure D-7 Image Monitor (Your actual monitor may appear different from picture.)
3.) Connect the video cable to the HP computer video connector and then route the cable to the
4-way video splitter.
The kit has the required cable for the connection between the computer and splitter input.
Secure the cable with supplied tie wraps.
4.) Route the BNC end of the video cable to the center row of BNC connectors on the video
splitter. Use caution when connecting BNC cables. Observe color codes and name on cables.
NOTICE
Potential for Equipment Damage. Touching the video signal cable connector pins may cause
them to bend. When connecting the video signal cable, check the alignment of the HD15
connector. Do not force the connector in the wrong way.
5.) From the kit, select the second short video cable (the one that has a BNC on one end and
HD15 on the other end) and connect this cable between the video splitter and the image
monitor.
6.) Add the video cables to the video splitter output side. There should be five BNC connectors.
a.) Ensure the cables are attached to the video inputs (from above).
b.) Attach the cables from the desktop image monitor to the video outputs on the left.
c.) Attach the cables from the remote monitor to the video outputs on the right.
Summary of Contents for Optima CT670