Chapter 1 - Position Subsystems
Page 93
sys
tems
2.) Run the main input power conductors and ground though flexible metal conduit (attached
between the PDU chassis and room duct-work) so you can move the PDU away from the wall
during service.
Figure 1-75 Flexible Conduit for PDU Power
3.) Locate the hole cover plate in Box 1 and attach the flexible metal conduit to the PDU.
4.) If present, remove the TS1 panel front cover.
5.) Strip the wires to fit securely on the power block.
6.) Observe incoming phases (L1, L2, and L3) and insert bare leads into power block.
7.) Insert vault ground into PDU vault ground lug.
8.) Tighten all fasteners securely and replace the TS3 front panel.
Figure 1-76 PDU Area Locations