Page 46
Section 7.0 - Gantry Bearing Gap Inspection
Section 7.0 Gantry Bearing Gap Inspection
All CT systems require a gantry bearing gap inspection before starting electrical calibration.
All international gantries are shipped in a wooden shipping crate that should not be removed until
it arrives at the installation site. This shipping container is designed to reduce the risk of shipping
damage.
The back cover needs to be removed to gain access to the gantry bearing.
7.1
Time and Personnel
7.2
Tools and Test Equipment
•
Standard tool kit
•
Inspection document
•
2.5 mm Allen wrench
•
Rear cover dollies (Qty = 2)
•
Flashlight
7.3
Preparation: Damage Indicators
If this is a Rotek bearing, a mark similar to that shown in
is visible on the inside edge
of the black-colored bearing assembly.
Figure 1-13 Gantry Bearing - Rotek Label
The mark has a serial number in the same format as:
ROTEK 2TF4-44E1B-MA91960-8F-2372-REV13.
Required Persons
Preliminary
Reqs
Procedure
Finalization
2 (FE or mechanical supplier)
15 min
15 min
5 min
Rotek Insignia
Gap