GE C
OMPANY
D
IRECTION
5472001-1EN, R
EVISION
6
O
PTIMA
CT680 S
ERIES
AND
O
PTIMA
CT670 I
NSTALLATION
M
ANUAL
Page 100
Section 1.0 - Introduction
•
Check for reasonable cable routing.
-
Take into consideration necessary take-up distances for equipment maintenance, etc.
-
Try to complete as neat a job as possible.
1.2
System Component Identification
Identify all system cables by the system component designators listed in
. Each end of a
system cable has a label, and may have a color near the connector, (refer to
) to indicate
the component and the jack identifier of the component. All cables are located on the lower right
shelf of the lean cart.
1.3
Cable Color Identifiers
The ends of the cables may be marked with a piece of blue, yellow, red, or orange colored tape to help
with the cable installation.
lists the subcomponent, and corresponding color.
DESIGNATOR
SYSTEM COMPONENT
CT2
Gantry
CT1
Patient Table
PDU
Power Distribution Unit
OC
Operator Console
(Console Computer)
WL
X-Ray ON Warning Light
A1
Primary Power Disconnect
SEO
System Emergency OFF
DS
Door Interlock Switch
BBNC
Broad-Band Network Connection
Table 2-1 System Component Identifiers
SUBCOMPONENT
COLOR
Gantry
Blue
Table
Yellow
PDU
Red
Console Computer
Orange
Table 2-2 Cable Color Identifiers
RUN
NO. DESCRIPTION
PART NUMBER
L
ONG
C
ABLES
(K
IT
P/N
5444556-2)
S
HORT
C
ABLES
(K
IT
P/N
5444556)
1 Facility MDP to Room Disconnect (A1)
cust. supplied
cust. supplied
2 Room Disconnect (A1) to PDU
cust. supplied
cust. supplied
3 Room Disconnect (A1) to System E-Off
cust. supplied
cust. supplied
Table 2-3 System Interconnect Cables