Revision E
MAC
5500 resting ECG analysis system
4-23
2020299-020
Troubleshooting: Frequently Asked Questions
Entering Patient Data
Q: Do I have to enter all the Information I see on the Patient Data screen?
A: In
System Setup
>
Basic System
>
Patient Questions
you can require that the
patient identification number, or medical record number) be entered. It is not a
requirement to enter any other data. However, we recommend that you enter the
patient name and identification number, at the least. If you are transmitting to
the MUSE system you will want to enter the
Location
number as well. If an
emergency situation dictates that you must complete the test. without entering
the patient data, make sure you edit the record to add the missing information
before you transmit it to the MUSE system.
Transmission
Losing Fields When Transmitting
Q: Why do I lose the Referring MD and Technician names off of my reports when
I transmit records to the MUSE system?
A: Your MAC 5500 may be transmitting to the SDLC modem on the MUSE
system instead of the CSI modem. Check in System Setup to make sure you are
transmitting to the MUSE system CSI phone number.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......