4-10
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Troubleshooting: Diagnostic Tests
Speaker Test
Use the arrow pad to select
Loud
or
Soft
. Press the
Enter
key to produce a loud or
soft tone. (The tone level difference is minimal.)
If tone is not audible, the main CPU board may need replacement. (Speaker is part
of CPU board.)
Highlight
Return
and press the
Enter
key to return to the
System Diagnostics
menu.
Keyboard Test
Press each key and verify that the key is highlighted on the screen and also displayed
at the top of the screen. (It is normal for a dim background image to remain on the
screen when you select the next key.) The numeric value that displays at the top of
the screen is the scan code representation of the pressed key.
NOTE
The display shows keys in the upper part of the screen that are only available
with the MAC 5500 ST keyboard.
Check both of the
Shift
keys by pressing each in combination with a letter to
display a capital letter.
Press the center of arrow pad control and verify that the word
IN
displays on
screen. Press
arrows
to change the displayed arrow position. A beep sounds
with each arrow press.
Press the
Shift
key and the
F6
key to exit the test.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......