Revision E
MAC
5500 resting ECG analysis system
3-15
2020299-020
Installation: MAC 5500 ST Requirements and Configuration
MAC 5500 ST Requirements and Configuration
Following is a list of interface requirements and setup configurations required for
the devices listed when used with the MAC 5500 ST option.
Compatible Blood Pressure Units
Colin - Model ST-780
Connection Requirements - Use cable PN 2008112-001 to connect from the
MAC5500 port
1
to the Colin serial port. Use cable PN 2008111-001 to connect
from the MAC 5500
ANA/TTL
port to the Colin QRS trigger input.
Device Configuration Requirements - None
MAC 5500 Configuration Requirements – At the
Main Menu
complete the
following in the order shown below:
Select
System Setup
,
Enter
System password
,
Exercise
Test
,
Inputs/Outputs
,
Change
Blood Pressure
to
Nipon-Colin
.
Sun Tech - Model Tango
Connection Requirements - Use cable PN 2008113-001 to connect from the MAC
5500 port
1
to the Sun Tech serial port. Use cable PN 2008111-001 to connect from
the MAC5500
ANA/TTL
port to the Sun Tech QRS trigger input.
Device Configuration Requirements – At the Tango Main Menu complete the
following in the order shown below:
Select
Utilities
,
Select
Device
,
Scroll to
ECG Trigger
and press enter,
Scroll to
DIGITAL
and press enter,
Scroll to
EXIT
and press
Enter
,
Scroll to
Test Parameters
and press
Enter
,
With
Technique
highlighted, press
Enter
,
Scroll to
DKA
and press
Enter
,
Scroll to
EXIT
and press
Enter
,
Scroll to
EXIT
and press
Enter
to return to the display screen.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......