4-12
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Troubleshooting: Diagnostic Tests
Roller Test
(Uneven darkness can appear if AC power is on during this test.)
After cueing, printing should start at approximately 13–14 mm on the page.
The pattern appears as diagonal light and dark wavy bands.
Isolated light spots indicate a flat spot on the roller and may indicate that
the print carriage assembly needs to be replaced.
A white line across the length of the page indicates a missing print head
dot.
Test Pattern II
A combination of Test Pattern I and Roller tests. The first three pages consist of a
series of triangular waveforms and various hashmarks. The fourth page is a partial
roller test.
Test Pattern II Continuous
Test Pattern II runs continuously until
Stop
is pressed.
Continuously Run Out Paper
This test is used in manufacturing to test how well the unit self-corrects tracking
problems.
62A
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......