C-6
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Recommended Separation Distances
The table below provides the recommended separation distances (in meters)
between portable and mobile RF communication equipment and the MAC 5500.
The MAC 5500 is intended for use in the electromagnetic environment on which
radiated RF disturbances are controlled. The customer or the user of the MAC 5500
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
MAC 5500 as recommended below, according to the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not listed above, the
recommended separation distance [
d
] in meters (m) can estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum
output power rating of the transmitter in watts (w) according to the transmitter
manufacturer.
NOTE
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people
Separation Distance in Meters (m) According to Frequency of Transmitter
Rated Maximum Output
Power of Transmitter in
Watts
150 kHz to 80 MHz
outside ISM bands
d = 1.2
150 kHz to 80 MHz
in ISM bands
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
P
P
P
P
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......