5-4
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Maintenance: Inspection and Cleaning
Inspection and Cleaning
Visual Inspection
Perform a visual inspection of all equipment and peripheral devices daily. Turn off
the unit and remove power before making an inspection or cleaning the unit.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Verify that all cords and connectors are securely seated.
Inspect keys and controls for proper operation.
Toggle keys should not stick in one position.
Knobs should rotate fully in both directions.
Exterior Cleaning
Clean the exterior surfaces monthly, or more frequently if needed.
1.
Use a clean, soft cloth and a mild dish washing detergent diluted in water.
2.
Wring the excess water from the cloth. Do not drip water or any liquid on the
equipment, and avoid contact with open vents, plugs, or connectors.
3.
Dry the surfaces with a clean cloth or paper towel.
Interior Cleaning
General
Check for dust buildup on the surfaces of the interior circuit boards, components,
and power supply. Use commercially available compressed air to blow away the
accumulated dust. Follow the manufacturers directions.
Thermal Printhead
Clean the thermal printhead every three months or more often with heavy use. A
build-up of thermal paper coating on the printhead can cause light or uneven
printing.
Use a solution containing alcohol on a nonwoven, nonabrasive cloth such as Techni-
Cloth to wipe off the printhead. Do not use paper toweling, as it can scratch the
printhead.
Summary of Contents for MAC 5500
Page 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Page 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Page 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Page 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Page 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Page 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Page 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Page 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Page 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Page 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Page 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Page 179: ......